The Emergency Gaseous Oxygen Inlet Station shall be an Amico Alert-1 series.
As required by NFPA 99, oxygen systems having the source of supply outside the building shall have an inlet to connect a temporary, auxiliary source of supply for emergency or maintenance situations. This connection shall include necessary valves and to allow for an emergency supply of oxygen and isolation of the normal source of supply. The inlet connection shall be a 1" female npt connection and be provided with a plug to prevent entry of foreign matter into the oxygen system when not in use.
A 1" brass ball valve controls access to the Emergency supply, and the gauge indicates the pressure being introduced into the oxygen pipeline. Maximum input pressure of 100 psi.
The Emergency Gaseous Oxygen Inlet shall be housed in a weather tight surface mount enclosure. The enclosure door shall be labeled 'Emergency Low Pressure Gaseous Oxygen Inlet' and shall be equipped with a staple for padlocking to allow entry only by
authorized personnel.
The interior of the enclosure shall be clearly labeled with instructions for connection and operation of the emergency oxygen inlet.
Bronze body check valves with female pipe threads on each end is required for installation in the main and emergency supply pipeline in accordance with NFPA 99.
A brass body relief valve with the relief pressure set @75 psi is required for installation in the emergency supply line. The relief valve shall be supplied factory cleaned for oxygen service and shall automatically reset after discharging to provide a positive seal.
All components will be sourced and cleaned for medical gas service. The sealing compound used will be compatible for medical gas service.
Amico products comply with NFPA 99.
The Emergency Gaseous Oxygen Inlet Station shall be an Amico Alert-1 series.
As required by NFPA 99, oxygen systems having the source of supply outside the building shall have an inlet to connect a temporary, auxiliary source of supply for emergency or maintenance situations. This connection shall include necessary valves and to allow for an emergency supply of oxygen and isolation of the normal source of supply. The inlet connection shall be a 1" female npt connection and be provided with a plug to prevent entry of foreign matter into the oxygen system when not in use.
A 1" brass ball valve controls access to the Emergency supply, and the gauge indicates the pressure being introduced into the oxygen pipeline. Maximum input pressure of 100 psi.
The Emergency Gaseous Oxygen Inlet shall be housed in a weather tight surface mount enclosure. The enclosure door shall be labeled 'Emergency Low Pressure Gaseous Oxygen Inlet' and shall be equipped with a staple for padlocking to allow entry only by
authorized personnel.
The interior of the enclosure shall be clearly labeled with instructions for connection and operation of the emergency oxygen inlet.
Bronze body check valves with female pipe threads on each end is required for installation in the main and emergency supply pipeline in accordance with NFPA 99.
A brass body relief valve with the relief pressure set @75 psi is required for installation in the emergency supply line. The relief valve shall be supplied factory cleaned for oxygen service and shall automatically reset after discharging to provide a positive seal.
All components will be sourced and cleaned for medical gas service. The sealing compound used will be compatible for medical gas service.
Amico products comply with NFPA 99.